Principle:
- Automatic sampling using a loop
- Automatic loop injection on packed column
- Isothermal gas chromatograph
- Detection of THT and TBM eluting from the column performed by MEDOR®
- Detector:Electrochemical wet cell Detector which is SSD Sulfur Specific Detector.
- Signal provided by reaction between the wet cell electrolyte and the sulfur compounds
KeyPoints:
- Fully compliant with ASTM D 7493-22 : Standard Test Method for Online Measurement of Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatograph and Electrochemical Detection.
- Automatic calibration/validation of the data, with embeded permeation tube
- Continuous monitoring with automatic online sampling
- Analytical performances:
- Specific, linear and very sensitive to sulfur compounds
- Results validation by automatic standard injection
- Long term stability using detector installed in reservoir
- Extremely low maintenance:
- Very long life time detector with electrolyte, up to 10 years
- Low gas consumption, can be reduced in option
- More than 10 years data storage : full chromatogram
- No cylinders required thanks to internal calibration tube and gas generators
- Automatic control with process device
- Intelligents system with tunable and interactive alarm levels
- Powerful VISTACHROM Chromatotec® software: Remote monitoring & injection control
- Full traceability with on board archiving of results and chromatograms
- QC Set up and control of threshold alarms
- Data export by MODBUS / 4-20 mA / 0-10 V
- Time stamp results
Options:
- Multiple stream selector (up to 32 streams with one analyser)
- Explosion proof version Exp or Exd for ATEX, IECEx zone 1 and 2 group IIC T4 and also for CSA C1D2 group B , C & D T4
- Calculation modules (Average / Statistics / Odor index and more)
- Automatic data transfert through: Module for 4 outputs 4-20 mA (with 0 mA for instrument default) / 0 – 10 V / MODBUS RTU or TCPI IP or MGS1
- Nitrogen or air generator for safe and hasardous area
- 24 V DC power supply
- Electric selection valve to reduce air consumption
- Liquid injection valve from LPG up to condensates
- High concentration injection valve up to 5%
- Wall mounted box XXX934
- Liquid injection valve for LPG analysis
- Touch screen on the front panel for Exp version (only ATEX)
- External cylinder inlet for automatic calibration from external cylinder
- MFC for multipoint calibration from internal CALIB and/or from cylinder
- Zero method