Principle:
The airmoCAL is designed for automatic validation of results. The calibration is ensured by permeation tubes (e.g. butane, hexane, benzene for airmOzone application) and/or by cylinder.
The permeation tubes swept by a constant pure air stream are installed in a precisely regulated temperature oven. These two parameters allow a full control of the emitted gas concentration. A certified cylinder can also be used for the calibration and validation of results.
For linearity, we use dilution points (e.g. If cylinder is 100 ppb BTEX: linearity curve will be: 0 /20 /40 /60 /80 /100 ppb (also for 1ppm) with 0 and 100 ppb corresponding to ZERO and Cylinder methods). We have in option:
- airmoCAL-D: the dilution in ensured by a critical orifice.
- airmoCAL-MFC: multiple dilution points, the dilution is ensured by two Mass Flow Controllers (MFC): to control air and cylinder flow input. These MFC allow doing a big range of dilution points.
The system is controlled by Vistachrom software and multiple methods can be programmed for the calibration: ZERO, Cylinder, dilution points, sample and CALIB
Multi-oven option:
CALIB: Sample generated by 3 permeation ovens. Each permeation oven contains up to 3 certified permeation tubes (e.g.: N-Butane, N-Hexane, Benzene, H2S, DMS…). Other permeation tubes are available upon request; a maximum of 6 tubes may be placed in one permeation oven. ZERO: Sample of dilution gas (Zero Air or pure N2) to verify the quality of the dilution gas or for analyzer cleaning.
The ovens were designed to be adapted to different permeation tubes size (ovens dimensions: 50 × 185 mm)
Multiplexer option:
This multiplexer is made with inert tubes and will be linked to different lines (one line with one METHOD,). The total number of methods is 16 (multiplexer + CALIB + Cylinder + … + …)
It allows to analyse on automatic sequence or to choose any of the different METHOD in manual mode.
It purges the different lines that are not in analysis to lower the sampling time for the next analysis and to avoid for any compound to be adsorbed in the line. To check the stability of the flow, a flow meter is installed on the front panel. It is important to control the flow every year with a flowmeter calibrated at ± 1%.
airmoCAL 200 option:
Specific with heated valve for semivolatile calibration. It includes heated sampling line from the airmoCAL to the analyzer.
VISTACHROM software:
Chromatotec® developed software system which enables:
- Remote control
- Full traceability with on board archiving of results
- Programation and control of threshold alarms
- Ponctual (laboratory) or continuous measurement
- Data export by MODBUS / JBUS / MGS1 / 4-20mA / 0-10V
Options:
- Gas purifier
- Module 4 x analog output 4-20 mA or 0-10 V.
- Alarms module • Automatic results validation.
- 24 V power supply.
- Zero Air or Nitrogen generator
- Sampling pump in 24V DC or airmoPUMP (XXX901 or XXX915)
- Pressure and temperature measure
- Syringe injection
- Automatic internal standard injection in all samples