airmo S

Online monitoring of Sulfur compounds by auto GC/FPD

  • Model: airmo S
  • Manufacturer: Chromatotec

Airmo S is a fully automatic isothermal gas Chromatograph for the analysis & monitoring of sulfur compounds with Sulfur Specific Detector. This solution is specific for application where low sensitivity on sulfurs must be reached. This auto GC uses a valve with 1 sample trap, and metallic capillary columns.

Targetted compounds:
    • Standard Version: CS2; Mercaptans (Methyl-SH / Ethyl-SH…); Sulfides (DMS / DMDS / DES…)
    • Upon Request: Isopropyl-SH / Terbutyl -SH
Application:
  • Ambient Air Monitoring
  • Waste Water Treatment Plant
  • Gas odorisation & deodorisation
  • Impurities in Pure Gas (CO2, CH4…)
Category:

Principle:

  • Trap temperature at +5° C and loop in option
  • All the pneumatic tubes are inert to Sulphur compounds
  • High efficiency dryer for sample water removal
  • Continuous sampling
  • Sample is injected into column oven for separation
  • Controlled oven selects optimum temperature based on application
  • Detection achieved by a sulfur specific dual flame photometric detector (FPD)
  • Calibration with internal permeation tube

Features:

  • The double flame ensures no quenching occurs
  • Automatic linearization of compounds
  • Identification of compounds is based on their retention time (Elution from the column)
  • Automatic data validation as standard with internal permeation tube (DMS)

VISTACHROM® software:

The software developed by Chromatotec® allows:

  • Remote monitoring
  • Full traceability through archiving of results and QC
  • Set up and control of threshold alarms
  • Export of data MODBUS / MGS1 / 4-20mA / 0-10V
  • Windows 10

Options:

  • Automatic data transfer through:
    • Modbus or MSG1 communication protocol
    • Analog 4-20mA or 0-10V output
  • Hydrogen Generator
  • Zero Air Generator
  • Multiple stream analysis with Multiplexer
  • Dilution system
  • Heated lines
  • N2 Generator
  • Dual mode chromaS/airmoS
Compounds analysed: Specific sulfur compounds analysis with trap (and loop in option): Me-SH, Et-SH, DMS, DES, DMDS, CS2 …
Options: n-propyl-SH, 2-butyl-SH, Isopropyl-SH, Terbutyl-SH Only with loop in option: H2S, SO2, COS
Detection limits with trap: • 0,1 ppb double Sulfur Compounds like DMDS, CS2 etc…
• 0,2 ppb single Sulfur Compounds (MM, DMS etc..)
Detection limits with loop in option: with loop = 250 µl and amplification 2
• 14 ppb for H2S/SO2/ MM/EM/DMS/DES/COS
• 8 ppb for DMDS & CS2
Detection range with trap (selectable ppm / ppb) • 0,2 to 10 ppb (amplification 2)
• 1 to 40 ppb (amplification 1)
Detection range with loop in option (selectable ppm / ppb) ppm or ppb : amplification 1&2
• 0,01 to 10 mg/m3 or 0 to 600 mg/m3
• % / ppm / ppb level depends on compounds
for ex H2S = 0,01 to 10 mg/m3 = 7 ppb to 7 ppm
Relative Standard Deviation: • RSD < 5% on concentration over 24hrs
• RSD < 1% on Retention Time over 24hrs
Supervisor: • Full result storage (data and chromatogram)
• Embedded computer Windows® based with LCD display
• 128 GB of Hardware storage on SSD memory
• 4 USB Connecting Port
• Two RS-232 ports
• Display: 10’’ TFT Color LCD
Cycle Times: • 10 min in airmoS mode or chromaS mode
• 30 min in dual mode (option): 10 min chromaS + 20 min airmoS
Gas Supply:
• Carrier Gas N2 recommanded / Air possible 3 bar (5ml/min) 1/8’’
• Sample Inlet
1 Bar or ambient pressure (airmopump)
• Sampling dryer:
150-200 ml/min of dry air
Detector
  • H2 at 2 Bar (70 ml/min) 1/8’’
  • Air at 3 bar (80 ml/min) 1/8’’
Pneumatic valve 90ml/commutation

Operation Temperature:

• Room with air conditioning: 10 to 25°c
Power Supply:

• Main: (230V / 50Hz) or (115V / 60Hz)
Electrical consumption:

• 250W (+50W for computer)
Dimensions and Weight:

• Rack: 482 mm (19’’)
• Height: 177,8 mm (4U) or 222 mmm (5U)
• Depth: 660mm
• Weight: 22 kg (With integrated computer)
To order:

Model:
airmo S (4U)

airmo S (5U)

C52000-Trap

C52022-Trap

TSP airmo S

Download

Product reviews

Review airmo S

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review airmo S
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    Related products

    ChromaTHC

    ChromaTHC Total Hydrocarbon Analysis System

    Direct Methane and Non-Methane Total Hydrocarbon (NMTHC) analysis by autoGC with Flame Ionisation Detection (FID)
    Find out more
    chromaPID : BTEX analyzer with PID detector

    chromaPID

    BTEX analyzer with PID detector
    Find out more
    ChromaFID

    ChromaFID

    Volatile Organic Compounds (VOC) analyzer
    Find out more
    airTOXICBTX PID

    AirTOXIC BTX PID

    a GC/PID for automatic monitoring of BTEX in air, water or soil
    Find out more
    AirmoVOC BTEX MCERTS

    AirmoVOC BTX MCERTS

    A GC/FID instrument for automatic, continuous monitoring of BTEX in air, water or soil
    Find out more
    DET H2O

    DET H2O

    ELECTROLYTIC MOISTURE MONITOR HYGROMETER
    Moisture monitoring in gaseous samples for concentration between 30 ppb and 5000 ppm
    Find out more
    airmoCAL

    airmoCAL

    Calibration device inside 19ˮ and 4U/5U rack
    Find out more
    Purge and Trap

    airmoPUMP

    a sampling pump
    Find out more

    Subscribe to Content

    If you need to copy the content, please leave your information, and we will respond as soon as possible. Sincerely, thank you!