HUMICAP Humidity and Temperature Probe HMP155

Highly accurate, reliable temperature and humidity data in even the harshest environments

Model: HMP155
Manufacturer: Vaisala – Finland
Features

  • Vaisala HUMICAPâ180R sensor: superior long-term stability
  • Optional warmed humidity probe and chemical purge
  • Plug-and-play
  • USB connection for service use
  • Use with DTR13 and DTR503 radiation shields and a Stevenson screen
  • Weather-proof housing IP66
  • Optional, fast temperature probe
  • Different output possibilities: voltage, RS-485, resistive Pt100
  • Applications: meteorology, aviation and road weather, instrumentation
Category:

Vaisala HUMICAP Humidity and Temperature Probe HMP155 provides reliable humidity and temperature measurement. It is designed especially for demanding outdoor applications.

Long-term stability

HMP155 uses the proven Vaisala HUMICAPâ180R sensor that has excellent stability and withstands well harsh environments. The probe structure is solid and the sensor is protected by default with a sintered teflon filter, which gives maximum protection against water, dust, and dirt.

Warmed probe and highhumidity environment

Measuring humidity reliably is challenging in environments where humidity is near saturation.

Measurements may be corrupted by fog, mist, rain, and heavy dew. A wet probe may not give an accurate measurement in the ambient air.

This is an environment to which Vaisala has designed this patented, warmed probe for reliable measurements. As the sensor head is warmed continuously, the humidity level inside it stays below the ambient level. Thus, it also reduces the risk of condensation forming on the probe.

Fast measurements

With its fast response time, the additional temperature probe for HMP155 is ideal for measuring in environments with changing temperatures. The membrane filter speeds up the relative humidity measurement.

Long lifetime

Protecting the sensor from precipitation, and scattered and direct solar radiation increases its lifetime. Thus, Vaisala recommends installing HMP155 in one of the following radiation shields: DTR503, DTR13, or Stevenson screen. For the additional temperature probe, an installation kit is available for Vaisala DTR502 Radiation Shield.

Calibration

The probe can be calibrated using a computer with a USB cable, with the push buttons, or with the MI70 indicator.

HMP155 humidity measurement performance

Sensor HUMICAP®R2, 180R and INTERCAP for typical applications HUMICAPâR2C, 180RC and INTERCAPC for applications with chemical purge and/or warmed probe
Observation range 0 … 100 %RH 
Response time at +20 °C (+68 °F) in still air with sintered Teflon filter 63 %: 20 s 

90 %: 60 s

Factory calibration uncertainty at +20 °C (+68 °F)1) ±0.6 %RH (0 … 40 %RH) 

±1.0 %RH (40 … 95 %RH)

Accuracy (including non-linearity, hysteresis, and repeatability)
At +15 … +25 °C (+59 … +77 °F) ±1 %RH (0 … 90 %RH) 

±1.7 %RH (90 … 100 %RH) 

At −20 … +40 °C (−4 … +104 °F) ±(1.0 + 0.008 × reading) %RH 
At −40 … −20 °C (−40 … −4 °F) ±(1.2 + 0.012 × reading) %RH
At +40 … +60 °C (+104 … +140 °F) ±(1.2 + 0.012 × reading) %RH
At −60 … −40 °C (−76 … −40 °F) ±(1.4 + 0.032 × reading) %RH 

 

1) Defined as ±2 standard deviation limits. Small variations possible (see also the calibration certificate).

 

HMP155 temperature measurement performance

 

Sensor Pt100 RTD element, Class F 0.1 IEC 60751
Observation range  −80 … +60 °C (−112 … +140 °F)
Response time for additional temperature probe in 3 m/s (7 mph) air flow  63 %: < 20 s 

90 %: < 35 s

Other measured variables Dew point / frost point temperature, wet bulb temperature, mixing ratio
Accuracy with voltage output
At −80 … +20 °C (−112 … +68 °F) ±(0.226 – 0.0028 × temperature) °C
At +20 … +60 °C (+68 … +140 °F) ±(0.055 + 0.0057 × temperature) °C
Accuracy with passive (resistive) output
According to Tolerance Class AA IEC 607511) ±(0.1 + 0.0017 × |temperature|) °C
Accuracy with RS‑485 output
At −80 … +20 °C (−112 … +68 °F) ±(0.176 – 0.0028 × temperature) °C 
At +20 … +60 °C (+68 … +140 °F) ±(0.07 + 0.0025 × temperature) °C

 

1) Tolerance Class AA IEC 60751 corresponds to IEC 751 1/3 Class B 

 

HMP155 operating environment

 

Operating temperature for humidity measurement  −80 … +60 °C (−112 … +140 °F)
Storage temperature −80 … +60 °C (−112 … +140 °F)
Operating humidity 0 … 100 %RH 
IP rating IP66

 

HMP155 inputs and outputs

 

Operating voltage 7 … 28 V DC
Minimum operating voltage 0 … 1 V output or RS‑485: 7 V 

0 … 5 V output, or warmed probe: 12 V 

0 … 10 V output, chemical purge, or XHEAT: 16 V

Outputs Voltage output: 0 … 1 V, 0 … 5 V, 0 … 10 V Resistive Pt100 4‑wire connection RS‑485
Average power consumption (+15 V DC, load 100 kΩ) 0 … 1 V output: < 3 mA 

0 … 10 V output: +0.5 mA 

RS-485: < 4 mA 

During chemical purge: Maximum 110 mA 

With warmed probe: Maximum 150 mA

Settling time at startup Voltage output: 2 s 

RS-485: 3 s 

 

HMP155 mechanical specifications

 

Dimensions (H × W) 279 × 40 mm (10.9 × 1.6 in)
Weight 93 g (3.25 oz) 
Length of additional T‑probe cable 2 m (6 ft 7 in)
Connection 8-pin male M12 connector 
Connection cables 3.5 m (11 ft 6 in), 10 m (32 ft 10 in), 30 m (98 ft 5 in)
Maximum wire size 0.129 mm2 (26 AWG)
Service cables USB connection cable

 MI70 connection cable

Materials
Filter Sintered Teflon or membrane
Housing Polycarbonate (PC)
Additional temperature probe Stainless steel AISI 316L
Cable PUR

HUMICAP® Humidity and Temperature Probe HMP155 Datasheet

Download

HUMICAP® Humidity and Temperature Probe HMP155

Download

Dependable weather measurements for harsh environments

Download

Product reviews

Review HUMICAP Humidity and Temperature Probe HMP155

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review HUMICAP Humidity and Temperature Probe HMP155
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    Related products

    Cảm biến nhiệt độ mặt đường Wintersense SDI-12

    Wintersense SDI-12 Non-Invasive Surface Temperature Sensor

    Forecast winter treatments for road surfaces, prevent traffic accidents and loss of life
    Find out more
    Cảm biến đo gió âm thanh 2D WINDSONIC1-L

    WINDSONIC4-L 2-D Sonic Wind Sensor

    with SDI-12 Output
    Find out more
    Cảm biến đo gió âm thanh 2D WINDSONIC1-L

    WINDSONIC1-L 2-D Sonic Wind Sensor

    with RS-232 Output
    Find out more
    Cảm biến nhiệt độ nước kỹ thuật số SDI-12 TempVue 50

    TempVue 50 SDI-12 Digital Water Temperature Sensor

    For years of scientific-grade water and soil temperature measurements
    Find out more
    Cảm biến nhiệt độ không khí kỹ thuật số Pt100 TempVue 20

    TempVue 20 Pt100 Digital Air Temperature Sensor

    Meets world meteorological organization recommendations
    Find out more
    Cảm biến nhiệt độ analog Pt100 TempVue 10

    TempVue 10 Pt100 Analog Temperature Sensor

    A market-leading, analog, ambient air temperature sensor
    Find out more
    Thiết bị đo mưa TE525WS-L

    TE525WS-L Rain Gage

    with 8 in. Orifice
    Find out more
    Thiết bị đo mưa TE525MM-L

    TE525MM-L Metric Rain Gage

    with 9.6 in. Orifice
    Find out more

    Subscribe to Content

    If you need to copy the content, please leave your information, and we will respond as soon as possible. Sincerely, thank you!