Barometric Pressure Transfer Standard PTB330TS

Portable, battery operated transfer standard with data capability

Model: PTB330TS
Manufacturer: Vaisala – Finland

Barometric Pressure Transfer Standard PTB330TS combines a PTB330 digital barometer with a hand-held MI70 indicator within a portable unit functioning as a transfer standard. The optional HMP155 probe is available for accurate humidity and temperature measurement. For professional meteorology, aviation, laboratories and demanding industrial applications.

Operational PTB330TS unit includes a PTB330 digital barometer, hand-held MI70 indicator, optional HMP155 humidity and temperature probe, optional MI70 Link PC software, a user’s guide and a weatherproof transfer case equipped with a shoulder strap.

Category:

Features

  • Portable, battery operated transfer standard with data logging capability
  • Barometric pressure with Vaisala BAROCAP® Digital Barometer PTB330
  • Excellent long term stability
  • Accurate measurement
  • Optional humidity and temperature measurements with HMP155
  • Vaisala HUMICAP®180R sensor – superior long-term stability
  • New, fast temperature probe
  • Chemical purge
  • Data can be logged, and transferred to a PC via MI70 Link software

Easy to use

Data can be logged, and transferred to a PC via MI70 Link software. Multilingual user interface with nine languages to select from.

Take the Lab to the Field

PTB330TS makes field inspection accurate and easy. It combines a PTB330 digital barometer with a hand-held MI70 indicator within a portable unit functioning as a transfer standard. The PTB330TS comes in a durable and weatherproof transport case that can be easily carried and shipped.

High measurement accuracy and excellent long-term stability

The PTB330 series features extremely high accuracy. Class A barometers for the most demanding applications are fine-adjusted and calibrated against a high-precision pressure calibrator. All the PTB330 barometers come with a NIST traceable, factory calibration certificate, also optional ISO/IEC 17025 accreditation calibration services are available.

General

 

Operating temperature
MI70 

PTB330

HMP155

−10 … +40 °C (+14 … +104 °F) 

−40 … +60 °C (−40 … +140 °F) 

−80 … +60 °C (−112 … +140 °F)

Operating humidity Non-condensing
Maximum pressure limit 5000 hPa abs
Power supply Rechargeable NiMH battery pack with AC-adapter or 4xAA-size alkalines, type IEC LR6
Menu languages English, Chinese, French, Spanish, German, Russian, Japanese, Swedish, Finnish 
Display LCD with backlight, graphic trend display of any parameter, character height up to 16 mm
Data logging capacity 2700 points
Alarm Audible alarm function
Compliance  • EMC Directive (2004/108/EC) Complies with the EMC product family standard EN61326‑1 

• Electrical equipment for measurement control and laboratory use 

• Basic immunity test requirements 

• Low Voltage Directive (2006/95/EC) 

• ROHS Directive (2002/95/EC)

Operation Time (Using Rechargeable Battery Pack) 
Continuous use with PTB330 11 h typical at +20 °C (+68 °F) 
Datalogging use  Up to 30 days

Measurement Performance

 

Barometric Pressure (PTB330)
Measurement range 500 … 1100 hPa
Linearity 1) ±0.05 hPa
Hysteresis 1)  ±0.03 hPa
Repeatability 1) ±0.03 hPa
Calibration uncertainty 2) ±0.07 hPa
Accuracy at +20 °C (+68 °F) 3) ±0.10 hPa
Temperature dependence 4) ±0.1 hPa
Total accuracy -40 … +60 °C (-40 … +140 °F) ±0.15 hPa
Long-term stability  ±0.1 hPa/year
Settling time at startup (one sensor) 4 s
Response time (one sensor) 2 s
Acceleration sensitivity Negligible
Relative Humidity (HMP155)
Measurement range 0 … 100 %RH 
Accuracy (incl. non-linearity, hysteresis and repeatability)
at +15 … +25 °C (+59 … +77 °F) ±1 %RH (0 … 90 %RH) 

±1.7 %RH (90 … 100 %RH)

at −10 … +40 °C (−4 … 104 °F) ±(1.0 + 0.008 reading) %RH
Factory calibration uncertainty at +20 °C (+68 °F ±0.6 %RH (0 … 40 %RH) 5) 

±1.0 %RH (40 … 97 %RH) 5)

Humidity sensor HUMICAP180R 

HUMICAP180RC

Response time at +20 °C (+68 °F) in still air with a sintered PTFE filter
63% < 20 s
90% < 35 s 

 

1) Defined as ±2 standard deviation limits of endpoint non-linearity, hysteresis, or repeatability error. 

2) Defined as ±2 standard deviation limits of inaccuracy of the working standard including traceability to NIST. 

3) Defined as the root sum of the squares (RSS) of endpoint non-linearity, hysteresis error, repeatability error, and calibration uncertainty at room temperature. 

4) Defined as ±2 standard deviation limits of temperature dependence over the operating temperature range. 

5) Defined as ±2 standard deviation limits. Small variations possible, see also calibration certificate

 

Thiết bị chuyển đổi áp suất khí quyển cơ bản PTB330TS

Download

Product reviews

Review Barometric Pressure Transfer Standard PTB330TS

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review Barometric Pressure Transfer Standard PTB330TS
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    Related products

    Skyspec instruments

    Passive remote sensing of atmospheric particles and gases
    Find out more
    OPEN PATH

    Open-path active remote sensing instrument

    Fast and accurate spectral trace gas measurements using active spectroscopy
    Find out more
    Hệ thống phân tích NO2-NOx-NO ICAD Series 210DL

    ICAD NO2-NOx-NO Analyzer SERIES 210DL

    Patented, fast, accurate and direct nitrogen dioxide detection
    Find out more
    Hệ thống phân tích ICAD-HONO/NO2-200L

    Icad HONO-NO2 Analyzer series 210

    Patented, fast, accurate and direct hono and no2 detection
    Find out more
    Bộ thu thập và xử lý dữ liệu Granite 10

    Granite 10 Measurement and Control Data-Acquisition System

    Ideal for automotive testing applications, featuring CAN bus and CAN FD output
    Find out more
    Mô-đun Đầu vào Cổng Dòng điện 100 Ohm CURS100

    CURS100 100-Ohm Current Terminal Input Module

    Expands capabilities, allows data logger to measure additional types of sensors
    Find out more
    Datalogger ba kênh cho cảm biến dây rung CRVW3-NE

    CRVW3-NE Three-Channel Vibrating Wire Datalogger without Enclosure

    Vibrating Wire Data Logger Using patented VSPECT™ technology
    Find out more
    Bộ Ghi Dữ Liệu Dây Rung Ba Kênh CRVW3

    CRVW3 Three-Channel Vibrating Wire Datalogger

    Self-Contained Vibrating Wire Data Logger Using patented VSPECT® technology
    Find out more

    Subscribe to Content

    If you need to copy the content, please leave your information, and we will respond as soon as possible. Sincerely, thank you!