H2S/TOS TS MEDOR

Sulfur compounds analyzer

  • Model: H2S/TOS TS MEDOR®
  • Manufacturer: Chromatotec

The H2S/TOS TS MEDOR® is an autoGC-ED (MEDOR® Electrochemical wet cell Detector) for the analysis and monitoring of H2S, TOS and TS by sum of H2S + TOS in natural gas and gaseous fuels.

Main applications:

  • Trace and ultra trace detection in Natural Gas / LPG / Gaseous Fuels
  • H2S in crude oil / Diesel / Fuel oil / Condensates / water
  • Deodorisation/Desulfurisation control ppb
  • Catalyzer protection
  • H2S in pure gas : H2 with special application
Category:

The H2S/TOS TS MEDOR® is an autoGC-ED (MEDOR® Electrochemical wet cell Detector) for the analysis and monitoring of H2S, TOS and TS by sum of H2S + TOS in natural gas and gaseous fuels.

Different configurations exist depending on the application and concentration range:

  • HS2/TOS TS MEDOR ppb for 0-1 ppm
  • H2S/TOS TS MEDOR ppm for 0-10 / 0-50 / 0-100 ppm

Principle

  • Automatic sampling using a loop
  • Automatic loop injection on metallic capillary column
  • Isothermal gas chromatograph
  • Detection of all compounds eluting from the column performed by MEDOR® detector : Electrochemical wet cell Detector which is a Sulfur Specific Detector (SSD)
  • Signal provided by electrochemical reaction between the wet cell electrolyte and the sulfur compounds
  • Fully compliant with ASTM D 7493-14 : Standard Test Method for On-line Measurement of Sulfur Compounds in Natural Gas and Gaseous Fuelsby Gas Chromatograph and Electrochemical Detection.
  • Compliant with ASTM D 5504-20 : Standard Test Method for Determination of Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and Chemiluminescence with alternative detector.

Key points

  • Continuous monitoring with automatic online sampling
  • Analytical performances :
    – Specific, linear and very sensitive to sulfur compounds
    – Results validation by automatic standard injection
    – Long term stability using detector installed in reservoir
  • Extremely low maintenance :
    – Very long life time detector with electrolyte, up to 10 years.
    – Low gas consumption, can be reduced in option
    – More than 10 years data storage
    – No cylinders required thanks to internal calibration tube and gas generators
  • Automatic control with process device
  • Intelligents system with tunable and interactive alarm levels
  • Powerful VISTACHROM Chromatotec® software :
    – Remote monitoring & injection control
    – Full traceability with on board archiving of results and chromatograms

Options

  • MODBUS RTU communication protocol
  • One modul for 4 x Analog output 4-20 mA or 0-10 V
  • 24 V power supply for transportable analyzers
  • Multiple stream selector (2 to 10).
  • airmopure (XXX031) or nitroxychrom (XXX913CS)
  • CALIB with DMS permeation tube
  • Explosion proof version Exp or Exd for ATEX, IECEx, Zone 1 and 2 and also for CSA C1D2
  • Internal electric heater and/or cooler for temperature regulation of the Exp/Exd cabinet with thermal insulation
  • Liquid sample system with purge for Sulfurs extraction from liquid phase (XXPurge ED Ex)
  • Multiple stream selector (2 to 32)
Compounds Analysed: • H2S, TOS (total of mercaptans and sulfides) and TS by sum (H2S + TOS)
Detection Limit: • H2S/TOS TS MEDOR® ppb: 5 ppb H2S (7 μg/m3)
• H2S/TOS TS MEDOR® ppm: 0.1 ppm H2S (0.1417 mg/m3)
Detection Range: H2S / TOS / TS:  • 0/10, 0/100 or 0/1000 (ppb or ppm)
• Low % with HC sampling valve
Relative Standard Deviation: • RSD < 3% on concentration over 48H.
• RSD < 0.6% on retention time over 48H.
Cycle Time: • RSD < 3% on concentration over 48H.
• RSD < 0.6% on retention time over 48H.
Cycle Time: • H2S / TS result in 2 min for 0-3 ppm range
• H2S / TS result in 5 min for higher range
Linearity: • > 0.995 for all compounds
Storage / Transfer of Results: • Hardware storage
• MODBUS communication protocol (optional)
• 4-20mA (optional)
Gas supply: • Carrier: Dry air or N2 (3 bars): 5 ml/min.
• CALIB: in continuous 50 ml/min. (option )
• CALIB during validation ~ 250ml/min (option)
• Pneumatic valve 90ml/commutation
Power supply: • Main (230V / 115V 50/60Hz)
• 24V battery (optional)
Electrical consumption: • 150 VA
Dimensions and weight: Rack: 4582 mm 19″
• Height: 222 mm (5U)
• Width: 482mm
• Depth: 600 mm
• Net weight: 20 Kgs
Wall mounted box:
• Height: 800 mm (1300 mm if XXPurge ED Ex is included)
• Width: 600 mm
• Depth: 300 mm
• Net weight: 40 Kg (50 Kg if XXPurge ED Ex is included)
To order: Model:
H2S/TOS TS MEDOR®

inbuilt computer – 5U (XX022)

M51022-TS
Rack or wall mounted

H2S / TOS TS MEDOR®

Download

Product reviews

Review H2S/TOS TS MEDOR

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review H2S/TOS TS MEDOR
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    Related products

    TDS LAB SERIES

    TDS LAB SERIES

    Advanced Thermal Desorption Spectrometry Systems
    Find out more
    máy quang trắc ngọn lửa xung

    Pulsed Flame Photometric Detector (PFPD)

    Pulsed Flame Photometric Detector
    Find out more
    Máy đo độ đục để bàn Turb 750 T

    Turb 750 T Benchtop Turbidimeter

    Simplify turbidity measurements without compromising data quality
    Find out more
    Hệ thống phân tích dòng chảy liên tục

    Flow Solution FS3700

    Continuous Flow Analyzer
    Find out more
    máy đo màu cod 910

    910 COD Colorimeter

    Single Parameter Colorimeter for COD
    Find out more
    máy đo màu clo 900

    900 Chlorine Colorimeter

    Single Parameter Colorimeter for Total and Free Chlorine
    Find out more
    máy đo màu cầm tay

    pHotoFlex

    pHotoFlex Colorimeters
    Find out more
    máy quang phổ 9800

    9800 Photometer

    Portable & Multiparameter Photometer for Water Quality Testing
    Find out more

    Subscribe to Content

    If you need to copy the content, please leave your information, and we will respond as soon as possible. Sincerely, thank you!