Principle:
Same calibration as XXX931 but outside the analyser.
The permeation tube, swept by a constant stream of pure gas (constant output which can be modified thanks to the pressure regulator), is installed in an oven regulated in temperature outside the analyzer and delivers a constant concentration of standard.
This feature allows validating all the analyses.
The gas of dilution used may be nitrogen or zero air (continuous output 50 ml / min and in dilution 300mL / min). In the sequence we program analyses and calibrations.
The electric gate of switching is included with a permeation tube (Benzene or DMS in standard).
A rotameter indicate the total dilution flow value used as working standard.
Example :
CALIB: Sample generated one permeation oven containing certified permeation tubes (e.g.: N-Butane, N-Hexane, Benzene, H2S, DMS…). One permeation tube is delivered in standard with the calibrator. Other permeation tubes are available upon request ; a maximum of 4 tubes may be placed in one permeation oven.
Ambient air: Sample from the analyzer (ambient air)