airmoVOC WMS

Monitoring of VOC in water

  • Model: airmoVOC WMS
  • Manufacturer: Chromatotec

Monitoring of VOC in water – BTEX included Based on EPA* 502.2 Method

Application:

  • Finished drinking water
  • Raw source water
  • Drinking water
  • Surface water
  • Wastewater (head space / ppt)
  • Rain water
  • Ambient air control
  • Urban/Non urban area pollution control
  • Indoor measurements
  • BTEX/PAMS/CE analysis
Category:

Principle

The airmoVOC WMS uses a valve with a sample trap.
It also features a metallic capillary column.
  • Miniaturization, sensitivity, mobility and flexibility are its main features.
  • Everything from the sample port up to the data storage is integrated in a wall mounted box.
  • Uninterrupted sampling with pre-concentration on absorbent tube
  • Gas chromatograph with metallic column with programmable temperature gradient oven.
Pressure control of the carrier gas by piezo-valve.
  • One week tested after production for quality control.
Vistachrom software enables the user to visualize and store data on a PC.
Furthermore it provides comfortable utilities to recalculate, calibrate and export data and to set-up measurement.
The software allows the calculation of retention time, area, mass or concentration profiles.
Purge: for on line analytic instrument
• Based on EPA 502.2 Method
• 5 ml of water sample as standard or optional 25 ml sparger
• Purge with inert gas: ultra pure N2 (Ultra High Purity) – 40 ml/min
• Sampling time: 11 minutes
• Inlet calibration is easy to perform thanks to a low sample volume required only 5ml of water
• Dead volume < 15 mL (volume between water and trap)
• Bubbles with a diameter < 3 mm at the origin of the frit
• Automatic rinse
Example of application
All VOC below can be analysed N° CAS
1,2-Dichloroethane 107-06-2
Benzene 71-43-2
1,2-Dichloropropane 78-87-5
Trichloroethylene 79-01.6
Toluene  108-88-3
Tetrachloroethylene  127-18-4
Chlorobenzene  108-90-7
Ethylbenzene 100-41-4
* m-Xylene 108-38-3
* p-Xylene  106-42-3
Styrene 100-42-5
All VOC below can be analysed N° CAS
o-Xylene 95-47-6
Isopropylbenzene 98-82-2
1,3,5-Trimethylbenzene 108-67-8
1,2,4-Trimethylbenzene 95-63-6  95-63-6
1,3-Dichlorobenzene 541-73-1
1,4-Dichlorobenzene 106-46-7
1,2-Dichlorobenzene 95-50-1
1,2,4-Trichlorobenzene 120-82-1
1,2,3-Trichlorobenzene 87-61-6
Hexachloro-1,3-butadiene 87-68-3
*SUM of M+P Xylene

Options

  • DET QMS for online GCMS
  • airmoVOC BTEX expert 1 ppt LDL
  • Automatic validation and calibration with internal CALIB
  • 24 V DC power supply
  • Integrated hydrogen and zero air generators for autonomous analysers
  • Multiplexer: 2 to 32 streams
  • 1 stream for water and 1 stream for air
  • Internal or external multipoint calibration and zero with CALIB MFC, XXXCYL MFC, airmoCAL PAH
  • airmoVOC C6C16 for more VOCs and S VOCs
  • Analog output 4-20 mA or 0-10 V

 

Analysis by airmoVOC:


 
3 main solutions : Up to 60 compounds with :

    • 60 compounds : our Purge & Trap 2 GC FID
    • 50 compounds : our Purge & Trap 1 GC FID
    • 50 compounds : our Purge & Trap 1 GC PID
  • BTEX and chlorine compounds
Detection limit:
  • < 0.001 µg/l for BTEX
Detection range:
  • 0.05 to 20 µg/l for surface water and finished drinking water
  • 0 to 50 000 µg/l for liquid and other liquid on demand
Relative standard deviation (RSD):
  • < 0.3% over 48h (Retention Time)
  • < 3% over 48h (Concentration)
  • < 10% for water analysis (Concentration)
Base Line: Zero drift:
  • < ±3%
Linearity:
  • R> 0.99 on all compounds
Supervisor:
  • Full result storage (data and chromatogram)
  • Embedded computer Windows® based with LCD display
  • 128 GB of Hardware storage on SSD memory
  • 4 USB Connecting Port
  • Two RS-232 ports
  • Display: 10’’ TFT Color LCD
  • MODBUS RTU / JBUS communication protocol
Cycle time:
  • 30 min or 60 min
Gas supply:
  • H2 (FID and carrier gas): 30 ml/min (supply 2 bar ; 1/16’’double ferrule
  • Air (FID) : 180 ml/min (supply 3 bar ; 1/8’’double ferrule
  • N2 (Purge): 40 ml/min (inlet 3 bars ; 1/8’’ double ferrule)
  • Sample inlet (vacuum pump) ; 1/4 double ferrule
  • Pneumatic valve 90ml/commutation
Operation Temperature:
  • Room with air conditioning: 10 to 25°c

Purge:

  • ZERO N2 analysis
  • ZERO WATER analysis (Blank)
  • Standard water analysis
Power supply:
  • Main: 230V / 50 Hz or 115V / 60 Hz
Electrical consumption:
  • Mean: 150 VA, Peak 360 VA
Installation in a wall-mounted cabinet:
  • Height: 1440mm
  • Width: 600mm
  • Depth: 300mm
  • Net weight: 80 Kg
To order: Model:
airmoVOC 624 A25022
airmoVOC-WMS A25022-502

 

TSP airmoVOC WMS

Download

Product reviews

Review airmoVOC WMS

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review airmoVOC WMS
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    Related products

    ChromaTHC

    ChromaTHC Total Hydrocarbon Analysis System

    Direct Methane and Non-Methane Total Hydrocarbon (NMTHC) analysis by autoGC with Flame Ionisation Detection (FID)
    Find out more
    chromaPID : BTEX analyzer with PID detector

    chromaPID

    BTEX analyzer with PID detector
    Find out more
    ChromaFID

    ChromaFID

    Volatile Organic Compounds (VOC) analyzer
    Find out more
    GAM 2000 Process Mass Spectrometer

    GAM 2000 Process Mass Spectrometer

    Multicomponent Online Gas Analyzer
    Find out more
    DET QMS

    DET QMS

    Mass spectrometer for continuous monitoring of gaseous compounds
    Find out more
    microVOC / microBTEX

    microVOC / microBTEX

    Portable VOC Analyzer
    Find out more
    microF : Portable Formaldehyde Analyzer

    microF

    Accurate, portable & user friendly Formaldehyde analyzer
    Find out more
    airmoVOC WMS

    Purge & Trap cabinet

    Surveillance of VOC in water - BTEX included Based on EPA* 502.2 Method
    Find out more