airmoMEDOR

Online analysis and monitoring of odorous sulfur compounds

  • Model: airmoMEDOR®
  • Manufacturer: Chromatotec

The airmoMEDOR® is an automatic gas chromatograph (autoGC) for the analysis and monitoring of trace amounts of Methyl-Mercaptan, Ethyl-Mercaptan, Di-Methyl-Sulfide, Di-Methyl-Di-Sulfide and Di-EthylSulfide by electrochemical wet cell detection (MEDOR® technology).

The airmoMEDOR® can be used for :

  • Monitoring of urban and non-urban pollution
  • Monitoring of industrial nuisance
  • Deodorization process
  • Fugitive emission
Category:

Principle:

  • Automatic sampling and concentration using an absorbant trap.
  • Desorption from the trap and injection into a metallic capilliary column.
  • Isothermal gas chromatograph.
  • The detection of all compounds eluting from the column is performed by an Electrochemical wet cell Detector which is Sulfur Specific Detector (SSD). This detector is using MEDOR technology.
  • Signal provided by reaction between the wet cell electrolyte and the sulfur compounds.

Key points:

  • Automatic calibration/validation of the data, with embedded permeation tube at ppb or ppt level on a daily basis in standard
  • Continuous monitoring with automatic online sampling
  • Analytical performances:
    • Specific, linear and very sensitive to sulfur compounds
    • Results validation by automatic standard injection
    • Long term stability using detector installed in reservoir
  • Extremely low maintenance
    • Very long life time detector with electrolyte, up to 10 years
    • Low gas consumption, can be reduced in option
    • More than 10 years data storage
    • No cylinders required thanks to internal calibration tube and gascgenerators
  • Automatic control with process device
  • Intelligent system with tunable and interactive alarm levels
  • Powerful VISTACHROM Chromatotec® software:
    • Remote monitoring & injection control
    • Full traceability with on board archiving of results and chromatograms
    • QC Set up and control of threshold alarms
    • Data export by MODBUS / 4-20 mA / 0-10 V
    • Time stamp results

Options:

  • Multiple stream selector (up to 48 streams with one analyzer)
  • Modeling software with weather station for live odor/chemical impact (XXXDISP)
  • Explosion proof version Exp for ATEX, IECEx zone 1 and 2 and also for CSA C1D2
  • Calculation modules (Average / Statistics / Odor unit OU e/m3 and more)
  • Automatic data transfer through: Module for 4 outputs 4-20mA / Modbus RTU or TCP IP / MGS1
  • Nitrogen or air generator
  • 24 V DC power supply
  • Electric selection valve to reduce air consumption
  • Wall mounted box XXX934
Sulfur compounds analysis: 5 compounds • In standard: MM, EM, DMS, DMDS and DES
• With loop option: H2S and SO2 in ppb range
• In option: IPM, TBM, NPM, MES, 2BM, IBM and NBM
Other compounds on request
Detection limits: • DES and DMS < 150 ppt (0.15 ppb)
• DMDS, MM and EM < 50 ppt (0.05 ppb)
• H2S and SO2 option < 5 ppb
Standard Detection Range: • 0 to 100 ppb
• Other range on demand
• Calculation: Odor Index, Total Sulfur, Total mercaptans
Relative Standard Deviation: • RSD < 3 % on concentration over 48H
• RSD < 0.5% on retention time over 48H
Cycle Time for the following different analyses:
• MM, EM, DMS, DMDS, DES 900s
• With H2S and SO2 option 2400s
With IPM, TBM, NPM, 2BM, IBM, NBM and MES option 1800s
Supervisor: • Embedded computer Windows® based with LCD display
• 128 GB of Hardware storage on SSD memory
Linearity: • > 0.995 for all compounds
Communication: • Ethernet, remote control
Gas supply: • Carrier: Dry air or N2 (3 bars): ≤ 4 ml / min.
• CALIB: in continuous ~ 50 ml / min.
• CALIB during validation ~ 250ml/min
• Pneumatic valve 90ml/commutation
Power supply: • Main: 230V 60Hz or 115V / 50Hz
• In case of power loss, the instrument will restart automatically
Electrical consumption: • 150 VA
Dimensions and weights: Rack: 19’’ (5U)

• Height: 222 mm

• Width: 482mm

• Depth: 660 mm

• Net weight: 22 Kg

To order: Model:
airmoMEDOR ppt – without computer – with peltier, dryer and CALIB

airmoMEDOR ppt – computer included – with peltier, dryer and CALIB

M53000 Trap

M53022 Trap

TSP airmoMEDOR

Download

Product reviews

Review airmoMEDOR

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review airmoMEDOR
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    Related products

    TDS LAB SERIES

    TDS LAB SERIES

    Advanced Thermal Desorption Spectrometry Systems
    Find out more
    máy quang trắc ngọn lửa xung

    Pulsed Flame Photometric Detector (PFPD)

    Pulsed Flame Photometric Detector
    Find out more
    Máy đo độ đục để bàn Turb 750 T

    Turb 750 T Benchtop Turbidimeter

    Simplify turbidity measurements without compromising data quality
    Find out more
    Hệ thống phân tích dòng chảy liên tục

    Flow Solution FS3700

    Continuous Flow Analyzer
    Find out more
    máy đo màu cod 910

    910 COD Colorimeter

    Single Parameter Colorimeter for COD
    Find out more
    máy đo màu clo 900

    900 Chlorine Colorimeter

    Single Parameter Colorimeter for Total and Free Chlorine
    Find out more
    máy đo màu cầm tay

    pHotoFlex

    pHotoFlex Colorimeters
    Find out more
    máy quang phổ 9800

    9800 Photometer

    Portable & Multiparameter Photometer for Water Quality Testing
    Find out more

    Subscribe to Content

    If you need to copy the content, please leave your information, and we will respond as soon as possible. Sincerely, thank you!