Total VOC with PID detector |
Sulfur compounds analysis: |
• Speciation and Total Reduced Sulfur by sum (TRS): H2S, SO2, DMS, MM, DES, DMDS, EM compounds |
Detection limits*: |
• Sulfur detector: less than 1 ppb for H2S
• TVOC PID detector: less than 10 ppb on DMS
• 50 UO E/m3 (limit of the olfactometer)
• *more sensitive analyzers available upon request with special application |
Range (adjustable depending on application): |
• 0/10 or 0/100 or 0/1000 (ppb or ppm)
• Calculation : Odor Index, TVOC, Total Sulfur, (OU e/m3 with XXXOU) |
Relative Standard Deviation for sulfur detector: |
• RSD < 3 % on concentration over 48H
• RSD < 0.5% on retention time over 48H |
Relative Standard Deviation for PID: |
• < 1% on concentration at 100ppb over 24h |
Cycle Time for the following different analyses: |
|
• H2S, SO2, MM, EM, DMS, DMDS, DES |
600s |
• With special application following compounds can be added :
IPM, TBM, NPM, 2BM, IBM, NBM, MES and THT |
1200s |
Supervisor: |
• Embeded computer Windows® based with LCD display
• 128 GB of Hardware storage on SSD memory |
Communication: |
• Ethernet, remote control |
Gas supply: |
• Carrier: Dry air or N2 (3 bars): ≤ 4 ml/min.
• CALIB: in continuous ~ 50 ml / min.
• CALIB during validation ~ 250ml/min for M52 configuration
• Pneumatic valve 90ml/commutation
• Sampling dryer: 150 to 200 ml/min |
Power supply: |
• Main: 230V 60Hz or 115V / 50Hz |
Electrical consumption: |
• 150 VA |
Dimensions and weights: |
|
Rack: 19’’ (5U)
• Height: 222 mm
• Width: 482mm
• Depth: 660 mm
• Net weight: 22 Kg |
Wall mounted box :
• Height: 1440 mm
• Width: 600mm
• Depth: 300 mm
• Net weight: From 45 to 70 Kg |
To order: |
Model: |
vigi e-nose (rack version)
vigi e-nose (wall mounted version)
NH3 analyzer associated (upon request)
NH3 sensor associated (upon request) |
V20022
V10022
D27000
XXXNH3 |