Aurora 1030W TOC Analyzer- OI Analytical/USA

This post is also available in: viVietnamese

The Aurora 1030W TOC Analyzer uses the proven heated persulfate wet oxidation technique to analyze organic contamination levels in liquid samples. The instrument can be programmed and calibrated to analyze samples containing organic carbon at levels as low as 10 ppb and as high as 30,000 ppm. In addition to its wide range, the Aurora 1030W has an extremely high daily throughput. Up to 300 samples can be analyzed in a 24-hour period (dependent on method and application).
The reaction chamber of the Aurora 1030W is thoroughly rinsed between analyses to eliminate residue from the previous sample. This ensures the Aurora 1030W maintains the low system background necessary for high sensitivity TOC measurements. This contrasts with combustion TOC analyzers in which residual salts accumulate on catalyst surfaces, degrading oxidation efficiency and causing higher blanks and background contamination.
Tested & Approved
The heated persulfate oxidation technique has been approved and adopted in numerous regulatory compliance methods, standards, and norms for water quality testing, including: Standard Method 5310C, USEPA 415.3, USP <643> / EU 2.2.44, ASTM D 4779 and 4839, USEPA-DBPR, USEPA-SPCC, ISO 8245, and EN 1484.

  • Technical Data
  • Accessories 
  • Applications 
Operating Principle Heated sodium persulfate oxidation
Measurement Range 10 ppb C – 30,000 ppm C (multiple calibration ranges or dilution required)
Instrument Detection LImit (IDL) 2 ppb C
Sample Injection Volume 10µL – 10mL
Method TC Acid and persulfate reaction
Method TIC Acidification with Phosphoric acid and sparging
Method TOC NPOC by heated persulfate oxidation or TC – TIC
Reagents Required Sodium persulfate, 5% phosphoric acid, rinsewater
Gas Supply N2 (99.998%), zero-grade air, or O2 (99.998%)

Bộ tiêm mẫu tự động 1088

 Made by: OI Analytical – USA

Bộ phân tích TOC trong mẫu rắn 1030S

 Made by: OI Analytical – USA

The Aurora 1030W TOC Analyzer is ideal for the following applications: drinking water, pharmaceutical cleaning validation, municipal wastewater, ground water/surface water, process water, boiler feed water and condensate, metal plating solutions, and ultrapure water.